Total Antioxidant Capacity (T-AOC) Assay Kit
| Cat.No. : | KITG-081 |
| Product Overview : | Total antioxidant capacity refers to the overall antioxidant level provided by various antioxidant substances and antioxidant enzymes, including antioxidant enzymes, vitamin C, vitamin E, and carotene. These components protect cells and organisms from oxidative stress damage caused by reactive oxygen species and free radicals. Measuring the total antioxidant capacity of plasma, saliva, urine and other body fluids, cell or tissue lysates, plant or traditional herbal extracts, and solutions of various antioxidants is important for a comprehensive and scientific evaluation of antioxidant capacity. This kit uses the FRAP (Ferric Reducing Ability of Plasma) method to determine total antioxidant capacity. Under acidic conditions, antioxidants reduce Fe³⁺-TPTZ to the blue Fe²⁺-TPTZ complex. The product has a characteristic absorption peak at 593 nm. The change in absorbance reflects the reducing capacity of the sample, namely its total antioxidant capacity. |
- Specification
- Gene Information
- Related Products
- Download
| Description : | Method: FRAP (Ferric Reducing Ability of Plasma), visible spectrophotometry. Pack Size: 70 tests / 50 samples. Storage Temperature: 2-8°C. For Research Use Only. Not for Use in Diagnostic Procedures. |
| Form : | Kit |
| Applications : | Quantitative determination of total antioxidant capacity in plasma, serum, saliva, urine and other body fluids; cell or tissue lysates; plant or traditional herbal extracts; and solutions of antioxidant substances. |
| Notes : | Instructions for Use Required Instruments and Reagents: Visible spectrophotometer, 1 mL glass cuvette with a 10 mm optical path, mortar/homogenizer, adjustable pipette, benchtop centrifuge, thermostatic water bath/incubator, concentrated sulfuric acid, and distilled water. 1. Sample Treatment (adjust the sample amount and ratio appropriately according to preliminary test results) 1) Tissue: Process the sample at a tissue mass (g) to Extraction Solution volume (mL) ratio of 1:(5-10). It is recommended to weigh 0.1 g tissue and add 1 mL pre-cooled Extraction Solution. Homogenize in an ice bath, centrifuge at 10,000 × g for 5 min at 4°C, collect the supernatant, and keep it on ice until measurement. 2) Bacteria or cells: Collect bacteria or cells by centrifugation into a centrifuge tube. Process the sample at a bacterial or cell number (10⁴) to Extraction Solution volume (mL) ratio of (500-1000):1. It is recommended to add 1 mL pre-cooled Extraction Solution to 5 million bacteria or cells. Disrupt the bacteria or cells by ultrasonication in an ice bath at 200 W, with 3 s sonication and 10 s intervals, repeated 30 times. Centrifuge at 10,000 × g for 5 min at 4°C, collect the supernatant, and keep it on ice until measurement. 3) Serum, plasma, saliva, urine, and other liquid samples: Heparin or sodium citrate may be used as the anticoagulant when preparing plasma; EDTA anticoagulation is not suitable. Centrifuge at 5,000 × g for 10 min at 4°C, collect the supernatant, and keep it on ice until measurement. Serum, saliva, or urine samples may be measured directly or stored at -80°C for no more than 30 days before measurement. 2. Assay Procedure 1) Preheat the spectrophotometer for at least 30 min, set the wavelength to 593 nm, and zero the instrument with distilled water. 2) Preheat the FRAP Working Solution at 37°C for 20 min before the assay. 3) Standard Curve Preparation: Dilute the 40 µmol/mL FeSO₄ standard solution with distilled water to prepare 0.1, 0.05, 0.025, 0.0125, 0.00625, and 0.003125 µmol/mL standard solutions. Standard 1: 500 µL of 0.1 µmol/mL standard solution + 500 µL Reagent 2; final Fe²⁺ concentration 0.05 µmol/mL. Standard 2: 500 µL of 0.05 µmol/mL standard solution + 500 µL Reagent 2; final Fe²⁺ concentration 0.025 µmol/mL. Standard 3: 500 µL of 0.025 µmol/mL standard solution + 500 µL Reagent 2; final Fe²⁺ concentration 0.0125 µmol/mL. Standard 4: 500 µL of 0.0125 µmol/mL standard solution + 500 µL Reagent 2; final Fe²⁺ concentration 0.00625 µmol/mL. Standard 5: 500 µL of 0.00625 µmol/mL standard solution + 500 µL Reagent 2; final Fe²⁺ concentration 0.003125 µmol/mL. Standard 6: 500 µL of 0.003125 µmol/mL standard solution + 500 µL Reagent 2; final Fe²⁺ concentration 0.00156 µmol/mL. Blank Tube 1: 500 µL distilled water + 500 µL Reagent 2. Blank Tube 1 only needs to be measured 1-2 times. Mix thoroughly, allow color development at room temperature for 10 min, and measure absorbance at 593 nm. Record the values as A_standard and A_blank1, and calculate ΔA_standard = A_standard - A_blank1. Use the final Fe²⁺ concentrations of 0.05, 0.025, 0.0125, 0.00625, 0.003125, and 0.00156 µmol/mL as the x-axis values and the corresponding ΔA_standard values as the y-axis values to plot the standard curve and obtain the linear regression equation y = kx + b. 4) Sample Measurement: Test Tube: 900 µL FRAP Working Solution + 30 µL test sample + 90 µL distilled water. Blank Tube 2: 900 µL FRAP Working Solution + 120 µL distilled water. Mix thoroughly and allow color development at room temperature for 10 min. Absorbance Measurement: Measure absorbance at 593 nm and record the values as A_test and A_blank2. Calculate ΔA_test = A_test - A_blank2. Substitute ΔA_test into the standard curve equation y = kx + b to calculate the sample concentration x in µmol/mL. Blank Tube 2 only needs to be measured 1-2 times. 3. Total Antioxidant Capacity Calculations Unit Definition: The total antioxidant capacity of a sample is expressed as the concentration of standard ions required to produce the same absorbance change (ΔA). 1) Based on tissue protein concentration: Total antioxidant capacity (µmol/mg prot) = (x × V_reaction)/(V_sample × Cpr) = 34 × x/Cpr. 2) Based on tissue sample mass: Total antioxidant capacity (µmol/g) = (x × V_reaction × V_sample_total)/(V_sample × W) = 34 × x/W. 3) Based on bacterial or cell number: Total antioxidant capacity (µmol/10⁴ cells) = (x × V_reaction × V_sample_total)/(V_sample × bacterial or cell count) = 34 × x/(bacterial or cell count). 4) Based on liquid sample volume: Total antioxidant capacity (µmol/mL) = (x × V_reaction)/V_sample = 34 × x. Where V_sample_total is the total volume of the test sample, 1 mL; V_reaction is the total reaction volume, 1.02 mL; V_sample is the volume of test sample added to the reaction system, 0.03 mL; W is the sample mass in g; Cpr is the sample protein concentration in mg/mL; and the bacterial or cell count is expressed in units of 10⁴. 4. Precautions 1) Reagent 2 is irritating. Avoid contact and take appropriate protective measures. Wear a laboratory coat and latex gloves during operation. 2) Avoid using samples that appear blue or nearly blue under acidic conditions, as they may interfere with the assay results. 3) Do not add detergents such as Tween or NP-40, or reducing agents that affect redox reactions, such as DTT or β-mercaptoethanol, to the samples. 4) If the measured absorbance is outside the linear absorbance range of the standards, dilute the test sample appropriately and repeat the measurement when the value is above the highest standard. When the value is below the lowest standard, appropriately increase the sample amount and repeat the measurement. Modify the calculation accordingly. 5) To ensure accurate results and avoid reagent loss, carefully read the instructions before measurement, confirm that reagent storage and preparation are adequate, verify that the operating steps are clear, and perform a preliminary test using 2-3 samples expected to show substantial differences. If any problem occurs during the procedure, seek technical assistance promptly. |
| Storage : | Kit Components and Storage: Extraction Solution: liquid, 60 mL × 1 bottle; store at 4°C. Pre-cool to 4°C before use. Reagent 1: liquid, 60 mL × 1 bottle; store at 4°C. Reagent 2: liquid, 20 mL × 1 bottle; store at 4°C protected from light. Reagent 3: liquid, 8 mL × 1 bottle; store at 4°C protected from light. Standard: powder, 1 vial containing 10 mg FeSO₄·7H₂O; store at 4°C. Before use, add 900 µL distilled water and 20 µL concentrated sulfuric acid to obtain a 40 µmol/mL FeSO₄ standard solution. |
| Reconstitution : | Reagent and Standard Preparation: 1. Standard Solution: Add 900 µL distilled water and 20 µL concentrated sulfuric acid to the vial containing 10 mg FeSO₄·7H₂O to obtain a 40 µmol/mL FeSO₄ standard solution. 2. FRAP Working Solution: Prepare immediately before use by mixing Reagent 1, Reagent 2, and Reagent 3 at a volume ratio of 10:1:1 according to the required amount. 3. Standard Dilutions: Dilute the 40 µmol/mL FeSO₄ standard solution with distilled water to prepare 0.1, 0.05, 0.025, 0.0125, 0.00625, and 0.003125 µmol/mL standard solutions. Required User-Supplied Reagent: Concentrated sulfuric acid (H₂SO₄, MW 98.078, CAS 7664-93-9). |
| Reaction conditions : | Preheat the FRAP Working Solution at 37°C for 20 min. For sample measurement, add 900 µL FRAP Working Solution, 30 µL test sample, and 90 µL distilled water to the test tube; add 900 µL FRAP Working Solution and 120 µL distilled water to Blank Tube 2. Mix thoroughly, allow color development at room temperature for 10 min, and measure absorbance at 593 nm. |
| Unit definition : | The total antioxidant capacity of a sample is expressed as the concentration of standard ions required to produce the same absorbance change (ΔA). |
Not For Human Consumption!